The theoretical
resolution of the light microscope was first defined by Abbe in the
following equation.
Abbe's
equation for theoretical resolution of the light microscope:

The actual resolution
achievable with a light microscope is not as great. We will discuss
the reasons for this later.
It is important
to understand and to recognize the various components of the light microscope.
The first and perhaps the most important element are the lenses.
Figure 2
The six simple lenses. a,b, & c are converging or positive
lenses. d,e, & f are concave or negative lenses.

Figure 3
Converging
and diverging lenses. Parallel rays enter a convex lens converge and
meet at the principle focus. The distance between the center of the
lens and the principal focus is the focal length. Parallel rays entering
a concave lens diverge and never meet.
The Objective Lens
is the first part of the imaging system; the objective lens forms a
primary, enlarged image of the object. Very fine details are distinguished
with the objective lens. The eyepiece sometimes called the ocular lens,
is the second lens, which forms a secondary, further enlarged image.
By multiplying the magnifying power of the objective lens and the magnifying
power of the ocular the final magnification is found. A Substage Condenser
lens is the third optical component. It is placed on a platform beneath
the object. Light is directed through the substage condenser and converges
to a point at the position of the specimen. The light rays diverge as
they pass through the specimen and form an inverted cone, whose base
is just large enough to fill the aperture of the objective. The size
of the light beam is controlled by a diaphragm beneath the condenser
called the aperture diaphragm.
The light source
should contain both a lens to project an image of the lamp filament
called a field condenser and a diaphragm to control the size of the
illuminated field called a field diaphragm.

Figure 4
Typical lamp for
light microscope.
Kohler Illumination
is the most common method of illumination. In Kohler illumination the
image of the source is projected by the field condenser onto the substage
condenser, to the top of the plane of the object. This method assures
even illumination.
Modern light microscopes
use several different modes of operation depending on the needs of the
investigator. The most common of these being brightfield microscopy
in which direct light passes through the objective aperture and illuminates
the background against which the image is seen. Since the structural
elements being resolved have little variance in refractive index, the
image will lack contrast and the details remain invisible. Small structure
detail can be revealed by changing the absorption of the object by means
of staining.
Many microscopes
have special lenses that allow an investigator to image using phase-contrast.
This method reveals details in specimens possessing very slight differences
in optical path or refractive index in the surrounding media. Phase
contrast requires no staining and can be used with living tissue. Contrast
is produced in this method by phase differences in the light leaving
the object to amplitude differences in the image. A phase plate alters
the optical path length traveled by diffracted light from that traveled
by direct light.
Figure 5
Light waves in
brightfield and phase contrast after passing through an object. Pathway
A represents the lightwave before encountering the object. Wave B represents
the wave phase after passage in brightfield (unstained mode). C compares
the wave phase of an object veiwed with phase contrast.
Differential interference
contrast (DIC) differs from phase contrast in that the image has a strong
relief and three-dimensional appearance. It must be remembered that
the impression of surface details are the results of the optics and
not the specimen for most biological samples. The optics for DIC consist
of a polarizer at the light source and Wollaston prisms in the condenser
and above the objectives. The beam passes through the polarizer, enters
the first prism where it is split in two. One beam vibration is parallel
to the prism and one is perpendicular. Both beams pass through the specimen
in parallel in close proximity and are recombined in the second prism.

Figure 6
Differential Interference Contrast Schematic.