Microscopes
1. Scanning Electron
Microscope
a. the electronics
in the SEM make a small probe, but stability is usually poor
b. the SEM
is suitable for a variety of attachments
c. sample
prep is minimal
d. 1-30
kV accelerating voltages are available
e. the resolution
is approximately 20-100Å
2. Microprobe
a. it may or may
not have high spatial imaging
b. its column
is configured to provide a high intensity, stable probe
c. it is
usually equipped with several wavelength spectrometer.
d. it is
intended for very accurate quantitative analysis
e. accelerating
voltages of 5-30 kV are typical.
3. Scanning Transmission
Electron Microscope
a. high resolution
is possible
b. phase
identification using electron diffraction imaging
c. EDS analysis
of very small areas
d. Voltages
of 75-300 kV are possible
e. specimen
prep is very difficult due to small specimen size
f. the operator
has access to all SEM signals as well
X-ray Instrumentation
1. Wavelength Dispersive
Spectrometry (WDS)
A small fraction
of the x-rays leaving the specimen strike a crystal and are reflected
onto the detector. The crystal will reflect only a very narrow range
of wavelengths of x-rays so Braggs law can be applied to determine
elemental content.
Since a particular
crystal will diffract only a very narrow wavelength of x-rays, one crystal
is useful for detecting only a few atomic numbers. Gypsum crystals will
detect atomic numbered elements 11 to 14 and sodium crystals will pick
up atomic numbered elements 16 to 37.
The focused x-rays
are detected when they pass through a thin plastic window and enter
a gas-filled cylinder containing a collector wire kept at a high positive
voltage. The x-rays cause an ionization of the argon-methane gas and
generate a flow of electrons to the wire. This current is measured and
quantified over time. The amount of current is directly related to the
energy of the original x-ray, so it is possible to determine the element
from which the x-ray was emitted.
2. Energy Dispersive
Spectroscopy (EDS)
A. The Detector
In EDS the detector
is usually made of a single silicon crystal, 2-3 mm thick. The crystal
is chosen for its purity and shape, but even so, conductivity varies
greatly and it is necessary to drift lithium into the crystal to fill
any vacant lattice spaces.
The atoms in this
crystal are covalently bonded. Each x-ray produces a photoelectron in
this valence band. A conduction band creating an electron-hole pair
dissipates the energy. Each electron-hole pair produces 3.8-3.9 eV.
Approximately 1,000 hole pairs are created per x-ray. The current generated
by the conducted charge is proportional to the energy of the absorbed
x-ray. There can be residual conductivity due to the random excitations
of electrons and cryogenic temperatures help to control this.
Because of their
extreme sensitivity, detectors must be protected from the interior of
the column. Most standard detectors use a beryllium window to achieve
this end. Lower energy x-rays (anything lighter than sodium) will not
pass through this window.
Next to the crystal
is a thin gold layer that aids in conduction. After this comes a space
termed the dead layer in which holes appear due to incomplete Li-drifting.
Some absorption occurs here, causing a tail on spectra peaks.
Many things can
happen as the x-ray enters the detector. It may not even reach the crystal
because the window, the gold layer or the dead layer may absorb it.
It may be too energetic and pass through the crystal with no interaction.
Or the interaction between the x-ray and the silicon atoms in the crystal
may be such that the silicon atom produces an x-ray of its own which
escapes from the detector and is ejected back into the column. This
last situation results in the formation of an escape peak. The formation
of an escape peak depends upon the angle of incidence into the detector
and the energy of the parent peak. The result of the escape is an artifactual
small peak 1.47 keV below the parent peak.
B. Preamplifier
or Field Effect Transistor
The signal is shuttled
here directly after it is collected in the crystal. The bias voltage
is subtracted from the pulses. A pulsed optical feedback is used to
reduce electronic noise and the result is a dead time in
which the system cannot take in any data as it processes a signal.
C. Pulse Processor/Amplifier
Pulses from the
preamp are shaped and amplified for analog to digital conversion. These
pulses must be separated because the height of the pulse carries information
about the energy of the x-ray. The pulse pileup rejecter may reject
pulses that come into the system together.
D. Analog to Digital
Converter and Multichannel Analyzer.
The pulses are converted
from analog to digital signals and then sorted into channels according
to height. The multichannel analyzer assembles the spectra and lets
you control it through the keyboard.

